1006 DC Millivolt Source

All about product
Specification
Reviews
0
Questions
0
Product Code: 1006
Reviews: 0
1 119.25€
Ex Tax: 1 119.25€
Specifications: (See all)
Additional options
Carry case (protective cover is removed when used). Rechargeable battery pack with mains charger. Calibration certificates: Traceable (factory) and accredited (ISO 17025)
Maximum Output Current
1 V and 100 mV ranges: 20 mA. 10 mV range: Up to short circuit value although it should be noted that loads of less than 1 kΩ will give greater than 0.1 % error.
Maximum Overload
The instrument can withstand continuous short circuit on the output for all ranges
Output
0 to 999.9 mV in 3 ranges: 1 V range: 0 to 999.9 mV in 0.1 mV steps 100 mV range: 0 to 99.99 mV in 10 μV steps 10 mV range: 0 to 9.999 mV in 1 μV steps
Output Noise Level
Less than 30 ppm of full scale
Description

The 1006 is an accurate millivolt source suitable for voltage injection applications. Three output ranges are provided to give adjustable output values from 1 μV to 1 V with a basic 0.02 % accuracy. 

For signal injection, the operator needs to switch on, check the battery condition, select the range, and set the required voltage using the thumbwheel switches. The instrument uses a precision reference diode and low temperature coefficient resistors to give a highly stable output.

Power is provided by 6 AA batteries. Battery life is several months, depending on usage. The battery condition is monitored by an indicator situated on the top of the unit.

The 1006 has up to 20 mA drive current and is short circuit and overload protected. An off/normal/reverse output polarity switch is provided.

Specifications
Main characteristics
Maximum Output Current
1 V and 100 mV ranges: 20 mA. 10 mV range: Up to short circuit value although it should be noted that loads of less than 1 kΩ will give greater than 0.1 % error.
Maximum Overload
The instrument can withstand continuous short circuit on the output for all ranges
Output
0 to 999.9 mV in 3 ranges: 1 V range: 0 to 999.9 mV in 0.1 mV steps 100 mV range: 0 to 99.99 mV in 10 μV steps 10 mV range: 0 to 9.999 mV in 1 μV steps
Output Noise Level
Less than 30 ppm of full scale
Output Polarity
Positive or negative switch selected. A centre ‘off’ position is also provided
Output Resistance
Less than 0.2 Ω on 1 V and 100 mV ranges. 1 Ω on 10 mV range
Output Voltage Stability
Less than 60 ppm/°C. Less than 100 ppm per 3 months (non-cumulative)
Reference Source
Precision bandgap reference diode, for stability and low temperature coefficient
Accuracy
± (0.02 % of setting + 0.02 % of range + 1 μV)
Autonomy
Power supply
6-AA size
Environmental parameters
Working temperature
–10 °C to 60 °C
Physical parameters
Size
H 200 x W 75 x D 110 mm
Weight
0.75 kg
Completeness
Additional options
Carry case (protective cover is removed when used). Rechargeable battery pack with mains charger. Calibration certificates: Traceable (factory) and accredited (ISO 17025)
Package contents
Protective cover (fitted), 6 x AA batteries (installed), and user manual