Main characteristics
2-wire variable resistance
1 Ω to 1.1 GΩ (± 100 ppm of setting)
4-wire decade resistance
1 Ω to 100 kΩ (± 30 ppm of setting)
Amplitude (Oscilloscope)
2 mV to 200 V (1 M Ω) and 1 mV to 2 V 50 Ω (Square-wave or DC) (± 0.05 %)
Capacitance
1nF, 10nF, 20nF, 50nF, 100nF, 200nF, 500nF, 1μF, 10μF, 20μF, 50μF, 100μF (± 0.2 % of setting)
Clamp meter adaptor
AC/DC Current up to 1000 A (DC, 45 to 90 Hz) (± 0.5 %)
Conductance
100 mS to 1 nS (decade values) (± 100 ppm of setting)
Current AC
10 μA to 22 A / 20 Hz to 5 kHz sine-wave 100 A with current transformer (option 9790) / 1000 A with clamp meter adaptor (option 9780) (±0.05 % of setting)
Current DC
0 to ± 22 A / 1000 A with clamp meter adaptor (option 9780) (± 60 ppm of setting)
Digital frequency
0.1 Hz to 10 MHz (± 20 ppm of setting)
Digital period
100 ns to 10 s (fixed values 1, 2, 5 sequence) (± 20 ppm of setting)
Duty cycle (Oscilloscope)
3 frequencies: 100 Hz, 1 kHz, 10 kHz, settable from 0 to 100 %
Fast-rise (Oscilloscope)
< 400 ps. Bandwidth checking up to 400 MHz
Frequency (Oscilloscope)
0.1 Hz to 10 MHz (± 0.1 ppm) / 20 MHz, 50 MHz, 100 MHz (± 20 ppm)
Fuse ratings
3.15 A anti-surge
Levelled Sine wave (Oscilloscope)
50 to 1000 kHz and 10 MHz to 2.2 GHz levelled sine-wave (0.35 to 3.5 V pk-pk) (Amplitude ± 0.5 %, Frequency ± 50 ppm)
Period (Oscilloscope)
10 s to 100 ns (± 0.1 ppm) / 50 ns, 20 ns, 10 ns (± 20 ppm)
Phase / Power factor
± 90.0° / 0.00 to 1.00 PF (± 0.25°)
Power
22 A, 1050 V, 23 kW, 40 to 500 Hz ACV: 0.05 %, DCV: 0.01 %. ACI: 0.1%, DCI: 0.05 %
RTD simulation
–180 to 850 °C. Type Pt100, Pt200, Pt500, Pt1000 (± 0.03 °C)
Thermocouple simulation
–210 to 2315 °C. Type J, K, R, T, S, B, E, N, C, L, U (± 0.05 °C)
Variable resistance
1 Ω to 1.1 GΩ (± 100 ppm of setting)
Voltage AC
2 mV to 1050 V / 20 Hz to 1 MHz sine-wave (± 0.05 % of setting)
Voltage DC
0 to ± 1050 V (± 15 ppm of setting)
Specifications
Warm up
30 minutes to full accuracy
Technical data
Interfaces
RS-232, GPIB and USB (via RS-232 adaptor). Command set: Standard SCPI
Power consumption
120 W typical, 200 W maximum
Completeness
Additional options
9762 Rubidium high stability frequency reference (for use with 5025E-S2/SCP models) - View data sheet /
9764 Current probe calibration adaptor (for use with 5025E-S2/SCP models) - View data sheet /
9773 Optical tacho adaptor - View data sheet /
9780 Clamp meter adaptor (1 and 50 turn coil) - View data sheet /
9790 100 Amp AC current transformer - View data sheet /
9760 Power amplifier (60 V AC, 90 V DC, 100 mA) - View data sheet /
9766 External low noise attenuator 1000:1 - View data sheet /
9767 External low noise attenuator 100:1 - View data sheet /
9085 Soft carry case /
9059 Hard transit case - View data sheet /
9728 19 " universal rack mount kit /
S2PTL Series 2 Premium test lead set (basic test lead set supplied as standard) /
EW01 Extended warranty - 3 years total covering parts and labour.
Package contents
Test lead set, PC virtual control software, RS-232 cable, RS-232 to USB adaptor, user manual