5030 Electrical Tester Calibrator

All about product
Specification
Reviews
0
Questions
0
Product Code: 5030
Reviews: 0
0.00€
Ex Tax: 0.00€
Specifications: (See all)
Additional options
C201 Traceable calibration certificate (Factory) / C137 Accredited calibration certificate (ISO 17025) / ECFLA EasyCal Full License (discounted price when purchased with 5030)
Current
6.000, 10.00, 30.00, 100.0, 300.0, 500.0, 1000 mA (±0.5 % of reading. ±1 % with x5 multiplier)
Current Multipliers
x0.5, x1, x2, x5
Interface
RS-232 and USB
Line Frequency Measurement
45.00 to 65.00 Hz (0.1 % of reading)
Description

The 5030 is a precision instrument designed to calibrate RCD, loop, insulation, earth testers and multifunction installation testers. It accurately simulates RCD trip times and measures currents produced by RCD testers. It replicates loop impedance and auto adjusts for local line impedance. It also provides insulation resistances and measures test voltages and currents.

User control is via the central navigation keypad. Scrolling though menus and settings is intuitive and easy, with measurements and settings shown on the LCD display. For loop and RCD tests the electrical tester connects directly to the dedicated front panel mains socket (regional type sockets available, specified on order). In these test conditions, if the device under test is faulty, the 5030 auto detects the fault, disconnects the output and warns the user. The 5030 is designed not to trip any RCDs on the local supply. For insulation and continuity, connection is via 4 mm terminals.

Software driven calibration can be performed by connecting the 5030 to a PC/laptop installed with EasyCal software.

Specifications
Main characteristics
Current
6.000, 10.00, 30.00, 100.0, 300.0, 500.0, 1000 mA (±0.5 % of reading. ±1 % with x5 multiplier)
Current Multipliers
x0.5, x1, x2, x5
Line Frequency Measurement
45.00 to 65.00 Hz (0.1 % of reading)
Line Voltage Measurement
100.0 to 260.0 V RMS (0.5 % of reading)
Local Loop Compensation
0 to 9.999 Ω (±0.5 % of value ± 30 mΩ)
Loop Impedance
1800, 330.0, 180.0, 33.00, 18.00, 3.300 (±0.5 % of displayed value ± 30 mΩ)
Phase Detection
0 ° or 180 °
Pre Trigger Delay
0 to 2000 ms
Pre Trigger Threshold
0 to 100 % of nominal current
Resistance
1 MΩ to 2000 MΩ 50 kΩ to 1990 kΩ (1% of value) / 0.1 Ω to 100.0 Ω 250 Ω, 500 Ω, 1.00 kΩ, 2.50 kΩ, 5.00 kΩ & 10.0 kΩ (1 % of value + 20 mΩ)
Resistor Values
1.800, 0.330, 0.150 and 0.050 Ω (±0.5 % of displayed value ± 30 mΩ)
Test Current
0 to 400 mA DC (0.5 % of range)
Test voltage
50.0 to 99.9 V DC 100 to 1200 V DC (1% of reading) / 0.0 to 50.0 V DC (0.5 % of range)
Trip Time
10 to 2000 ms (±0.5 ms)
Waveforms
AC, DC, half wave rectified
Max Current
2500 mA
Specifications
Warm up
30 min
Technical data
Power supply
220 to 240 V AC 50 Hz
Interface
Interface
RS-232 and USB
Environmental parameters
Working temperature
10 to 35 °C
Storage temperature
-10 °C to 50 °C
Working humidity
< 80% non condensing
Physical parameters
Size
W 430 x H 155 x D 255 mm
Weight
8 kg
Completeness
Additional options
C201 Traceable calibration certificate (Factory) / C137 Accredited calibration certificate (ISO 17025) / ECFLA EasyCal Full License (discounted price when purchased with 5030)
Package contents
User manual, RS-232 cable, USB adaptor/cable, power lead
Reviews
0 / 5
average product rating
0
0
0
0
0

There are no reviews for this product, be the first to leave your review.