Calibration Documentation
Report of calibration with data included; 17025 accredited report optional
Voltage DC
0 to ±1020V (Best 1 Year Specification: ±0.01% of setting)
Warm up Time
Twice the time since last warmed up, to a maximum of 30 minutes
Current DC
0 to ±20.5A (Best 1 Year Specification: ±0.05% of setting)
Settling Time
Less than 5 seconds for all functions and ranges except as noted
Description

Accurate, reliable analog meter calibration

The 5080A Multi-Product Calibrator calibrates your analog workload accurately and reliably, thanks to its high voltage and current compliance. With maximum burden up to 800 mA for ac/dc voltage, and voltage up to 50 V for ac/dc current, 5080A calibrators can drive a wide range of analog meters.

Options and accessories expand workload coverage

Options and accessories enable you to use the 5080A Multi-Product Calibrator to calibrate an even broader workload, including:

  • Clamp meters. The 9100-200 10/50 turn coil and 5500A/COIL 50-turn current coil enables the 5080A to calibrate most popular clamp meters at currents up to 1000 A rms amps.
  • Oscilloscopes. Oscilloscope calibration to 200 MHz quickly, easily, and cost effectively. Verify dynamic response, bandwidth, timing, multiple triggering functions, and input resistance.
  • Megohm meters. This option sources high ohms, high voltage resistors up to 18 G ohms. It also measures high voltage outputs.

Protective circuits prevent damage

Mains voltage inadvertently applied to a calibrator’s output terminals can cause extensive damage, requiring costly repairs. Electrical protection for calibrator outputs is vital for daily operation. The 5080A calibrator’s innovative protection circuitry prevents it from being damaged by reversed input voltage, so you can use it with confidence day after day.

Highest voltage and current compliance

The 5080A has the highest voltage and current compliance of any calibrator in the Fluke Calibration multi-product and multifunction calibrator families, making it an ideal solution for calibrating analog meters and other instruments requiring higher drive capability for proper operation.

Maximum burden or compliance voltage
Functions
ModelDC voltageAC voltageDC currentAC current*
5080A600 mA800 mA50 V44 V
9100 20 mA20 mA4 V4 V
5500A 10 mA10 mA4.5 V3 V
5520A 10 mA10 mA7 V5 V
5700A 50 mA50 mA10 V7 V
5720A 50 mA50 mA10 V7 V

* With AC LCOMP on

Automate the 5080A Multi-Product Calibrator to increase efficiency and throughput

For the full spectrum of calibration automation and asset management, choose MET/CAL software

The solutions you need, from the leader in calibration

Fluke Calibration is a premium brand, well known around the world for its accurate, dependable, high quality products. Long known as a leader in dc and low frequency ac calibration, Fluke Calibration is also recognized for its offerings in temperature, pressure, power, process, and rf calibration. Fluke Calibration provides the calibrators, standards, software, service, support and training you need for a complete solution in your cal lab.

Specifications
Technical Specifications
Voltage DC
0 to ±1020V (Best 1 Year Specification: ±0.01% of setting)
Current DC
0 to ±20.5A (Best 1 Year Specification: ±0.05% of setting)
Resistance
0 to 190 MO, 19 values in x1 and x1.9 (Best 1 Year Specification: ±0.025% of setting)
Voltage AC
1 mV to 1020V (45Hz to 1 kHz, Sine Best 1 Year Specification: ±0.1% of setting)
Current AC
29 mA to 20.5A (45 Hz to 1 kHz, Sine Best 1 Year Specification: ±0.1% of setting)
DC Power
10.9 mW to 20.5 kW (Best 1 Year Specification: ±0.11% of setting)
AC Power
10.9 mW to 20.5 kW (45-65 Hz, PF=1 Best 1 Year Specification: ±0.36% of setting)
Phase
0 to ± 179.99° (45 Hz to 1 kHz Best 1 Year Specification: ±0.25° of setting)
Frequency
45 Hz to 1 kHz (Best 1 Year Specification: ±50 ppm of setting)
General Specifications
Warm up Time
Twice the time since last warmed up, to a maximum of 30 minutes
Settling Time
Less than 5 seconds for all functions and ranges except as noted
Standard Interfaces
RS-232 and Ethernet
Temperature Performance
Operating: 0°C to 50°C / Calibration (tcal):15°C to 35°C / Storage:-20°C to +70°C
Temperature Coefficient
Temperature coefficient for temperatures outside tcal ±5 °C is 10 % of the stated specification per °C for temperatures in the range of 0 °C to 35 °C. Above 35 °C, the temperature coefficient is 20 % of the stated specification per °C
Relative Humidity
Operating: <80% to 30°C, <70% to 40°C, <40% to 50°C / Storage:<95%, non-condensing
Altitude
Operating: 2,000m (6,500 ft) maximum / Non-operating: 12,200m (40,000 ft) maximum
Safety
Meets EN. 61010-1:2001, CAN/CSA-C22.2 No. 61010-1-04, UL 61010-1: 2004
Analog Low Isolation
20V
EMC
Meets EN613261:2006
Line Power
Line Voltage (selectable): 100V, 120V, 220V, 240V / Line Frequency: 47 to 63 Hz / Line Voltage Variation: + 10% about line voltage setting
Power Consumption
600 VA
Dimensions
H 19.3 cm W 43.2 cm ( 44.3 cm including handles) D 53.8 cm
Weight (without options)
22 kg
Absolute Uncertainty Definition
The specifications include stability, temperature coefficient, linearity, line and load regulation, and the traceability of the external standards used for calibration. It is not necessary to add anything to determine the total specification for the temperature range indicated
Specifications confidence level
99%
Related documentation
Calibration Documentation
Report of calibration with data included; 17025 accredited report optional
Reviews
0 / 5
average product rating
0
0
0
0
0

There are no reviews for this product, be the first to leave your review.

Answers & questions

No questions about this product, be the first and ask your question.

Fluke 5080A High Compliance Multi-Product Calibrator
0.00€
Модули для Opencart