Quantity
|
Cost
|
||
|
|
|||
|
|
|||
Personal Gamma Radiation Dosimeter EcotestCARD DKG-21
Downloads:
PDF Instruction 1
The EcotestCARD DKG-21 is a compact and highly reliable personal dosimeter designed for continuous monitoring of individual dose equivalent (DE) and dose equivalent rate (DER) of gamma and X-ray radiation. Built around an energy-compensated Geiger-Muller counter, it delivers exceptional measurement accuracy across a broad energy range from 0.05 MeV to 6 MeV. Its proven field performance was demonstrated during emergency operations at the Fukushima NPP, proving its durability under harsh environments.
Equipped with an integrated non-volatile memory, the device records dose accumulation history linked to a real-time clock. Data can be seamlessly transferred to a PC via an infrared interface (IrDA) for automated processing, generating analytical reports, and long-term archiving. The instrument supports independent threshold programming for both dose rate and accumulated dose, triggering visual and audible alarms upon limit breaches.
With an ultralight credit-card form factor weighing only 80 grams, an IP31 enclosure, and up to 3,000 hours of continuous operation from a rechargeable Li-ion battery, the EcotestCARD is the premier choice for professional B2B individual dosimetry systems.
Applications
- Nuclear power plants and nuclear research facilities
- Medical radiology, radiotherapy, and diagnostic imaging centers
- Industrial radiography and non-destructive testing (NDT)
- First responders, civil defense, and emergency services
- Customs, border protection, and hazardous material monitoring
- Radiological laboratories and industrial dosimetry systems
Specifications
| Parameter | Value |
| Measurement range of DER (Dose Equivalent Rate) | 0.1 uSv/h to 1 Sv/h |
| Measurement range of DE (Individual Dose Equivalent) | 0.001 mSv to 9999 mSv |
| Energy range of registered gamma radiation | 0.05 MeV to 6 MeV |
| Energy dependence (0.05 to 1.25 MeV relative to 0.662 MeV) | Within +-25 percent |
| Basic relative error of DER (1 uSv/h to 10 uSv/h) | +-20 percent |
| Basic relative error of DER (10 uSv/h to 1 Sv/h) | +-15 percent |
| Basic relative error of DE (0.01 mSv to 9999 mSv) | +-15 percent |
| Response time for DER (1 uSv/h to 10 uSv/h) | Max 10 s |
| Response time for DER (10 uSv/h to 1 Sv/h) | Max 5 s |
| Continuous operation time on full charge | Min 3000 hours |
| Power supply | Integrated rechargeable Li-ion battery 3.7 V |
| Protection rating | IP31 |
| Dimensions | 90 x 55 x 10 mm |
| Weight | 0.08 kg |
Package Contents
- EcotestCARD DKG-21 dosimeter
- Integrated Li-ion rechargeable battery
- Micro-USB battery charger (5V, 0.5A)
- Protective leather case
- Operating manual
- USB/IrDA infrared PC adapter (optional upon order)
- Specialized PC software for Windows (optional upon order)
- Packaging box