Quantity
|
Cost
|
||
|
|
|||
|
|
|||
The CTi 4M high-speed infrared thermometer features an exceptionally fast exposure time of 110 μs, making it ideally suited for precise temperature measurement in high-speed or rapidly changing processes. Due to its unique mid-wavelength sensitivity and a temperature range from 0 °C to 500 °C, the pyrometer optris CTi 4M is suitable for measurements across various materials, like metals, metal oxides, ceramics or any material with unknown or changing emissivity. The pyrometer is a great alternative to LT sensors for high speed and low-temperature applications of challenging materials. The electronics box of the measuring device allows a flexible terminal device through selectable analog outputs as well as various optional digital interfaces.
- Temperature range: 0 °C to 500 °C
- Spectral range: 2.2 µm - 6.0 µm
- Response time: 320 μs
| MODEL | CTi 4ML |
| DETECTOR | |
| Detector | InAsSb |
| Measurement Principle | Single Color |
| Spectral Range | 2.2 – 6 μm |
| Response Time | 300 µs |
| Exposure Time | 110 µs |
| Sampling Frequency | 3300 Hz |
| Exchangable Sensing Head | No |
| OPTICAL | |
| Distance to Spot ratio (D:S) | 10:1 |
| Spot size (SF optics) | 40 mm at 400mm |
| Other optics (FF/CF optics / add. CF lens) | CF: 5.0 mm at 50 mm |
| Focus | fixed |
| SIGHTING | |
| Laser Sighting | none |
| Video Sighting | none |
| MEASUREMENT | |
| Object Measurement Range | 0 °C … 500 °C |
| Accuracy | ±0.3 % of reading +2 °C **2) |
| Repeatability | ±0.16 K |
| Temperature Coefficient | ±0,05 K/ K or ±0,03 %/ K **3) |
| Thermal Sensitivity (NETD) | 70 mK |
| Warm-up time | none |
| Emissivity /Transmissivity/ Reflectivity | 0.05 – 1.100 |
| Slope | N/A |
| INTERFACES | |
| Interface | USB / RS232 / RS485 / Ethernet / Relay |
| Supported Protocols | USB 2.0 optional: EtherNet/IP / Ethernet TCP / Modbus TCP / Modbus RTU / Profinet |
| Compatible Software | CompactPlus Connect, IRmobile |
| ANALOG INPUT/OUTPUT | |
| Direct output/input | 0-20 mA/4-20 mA/0-5 V/0-10 V/t/c (K/J) |
| Output impedances | max. 500 Ω (with 8-36 V DC) for mA output min. 100 kΩ load impedance for mV output thermocouple 20 Ω |
| Cable length | 3 m (standard), 8 m, 15 m |
| SIGNAL PROCESSING | |
| Configuration | CompactPlus Connect or IRmobile |
| Operation | Autonomous Standalone |
| Capabilities | Average, Peak hold, Valley hold, Advanced peak hold with threshold and hysteresis, Triggered signal output, Triggered peak hold function (adjustable via software) |
| GENERAL | |
| Size | Sensing head: Ø12mm x 35 mm, M12x1 Electronics: 122 mm x 70 mm x 27 mm |
| Housing Material | Stainless Steel |
| Weight | 40 g (sensing head) / 420 g (electronics) |
| Country of Origin | Germany |
| ENVIRONMENTAL & CERTIFICATIONS | |
| Operating Temperature Range | 0 °C … 70 °C (sensing head) 0 °C … 70 °C (electronics) |
| Storage Temperature Range | –40 °C … 85 °C (sensing head) –40 °C … 85 °C (electronics) |
| Relative humidity | 10 – 95 %, non condensing |
| Protection Class | IP 65, NEMA 4 |
| EMC | 2014/30/EU |
| Shock | IEC 60068-2-27 (25 G and 50 G) |
| Vibration | IEC 60068-2-6 (sinus shaped) IEC 60068-2-64 (broadband noise) |
| Standards | CE, UKCA, RoHS |
| POWER | |
| Power Supply | 8 – 30 V DC / 5 V USB / max. 1.2 W |
| Power Consumption | <50mW |
| Part number | OPTCTI4ML |
| Additional Remarks | 1) 90% energy 2) at ambient temperature 23 ± 5°C, object temperature > 0°C ; 3) for ambient temperatures <18°C and >28°C; whichever is greater |