Personal Electronic Dosimeter ECOTEST EPD-27 (DoseG / DoseGX)

Downloads:
PDF Instruction 1

The ECOTEST EPD-27 (available in DoseG and DoseGX modifications) represents a new generation of personal electronic dosimeters engineered for precise measurement of personal dose equivalent Hp(10) and personal dose equivalent rate Hp(10) of gamma and X-ray radiation. Built around an advanced scintillator detector with a Silicon Photomultiplier (SiPM), the EPD-27 provides outstanding sensitivity, high accuracy, and fast reaction times in dynamic radiation fields.

Designed for both autonomous operation and integration into automated personal dosimetry control systems (PDCS), the instrument features non-volatile memory for event history and dose accumulation tracking. The DoseG model focuses on gamma radiation detection (0.05 to 10 MeV), while the DoseGX model extends capabilities to soft X-ray radiation (12 to 200 keV). Housed in a glass-filled shock-resistant plastic casing rated IP67, the dosimeter withstands severe mechanical impacts, drop tests, and harsh environmental operating conditions.

Applications

  • Nuclear power plants and nuclear fuel cycle industrial facilities
  • Medical physics, radiology, and X-ray diagnostic departments
  • First responders, hazmat teams, and civil defense agencies
  • Industrial radiography, NDT testing, and isotope laboratories
  • Automated personnel radiation dose tracking and ALARA monitoring

Specifications

ParameterDoseG ModificationDoseGX Modification
Gamma Radiation Dose Equivalent (IED) Range1.0·10⁻⁷ to 10 Sv1.0·10⁻⁷ to 10 Sv
Gamma Radiation Dose Equivalent Rate (PIED) Range1.0·10⁻⁶ to 10 Sv/h1.0·10⁻⁶ to 10 Sv/h
X-ray Radiation Dose Equivalent Rate (PIED) RangeN/A1.0·10⁻⁶ to 1.0·10⁻¹ Sv/h
X-ray Energy RangeN/A12 to 200 keV
Gamma Energy Range0.05 to 10 MeV0.05 to 10 MeV
Basic Relative Error (137Cs)±15% (Dose), ±15% to ±20% (Dose Rate)±15% (Dose), ±15% to ±20% (Dose Rate)
Detector TypeScintillator with SiPMScintillator with SiPM
Data Transfer / PC InterfaceInfrared Port (IrDA)Infrared Port (IrDA)
AlarmsAudio, Visual (LED), VibrationAudio, Visual (LED), Vibration
Ingress Protection & Shock ResistanceIP67, 6 drops from 1 m, 60 shocks (10 cm drop equiv)IP67, 6 drops from 1 m, 60 shocks (10 cm drop equiv)
Power Supply & Operating Time3.7 V Li-Po 400 mAh; up to 170 h continuous3.7 V Li-Po 400 mAh; up to 170 h continuous
Operating Temperature & Humidity-20°C to +50°C; up to 95% at 35°C-20°C to +50°C; up to 95% at 35°C
Dimensions & Weight84.5 x 55 x 24.5 mm; 0.11 kg84.5 x 55 x 24.5 mm; 0.11 kg
OIML D 11:2018 Environmental ClassE2 (Electromagnetic), M1 (Mechanical)E2 (Electromagnetic), M1 (Mechanical)